The room or lab designated for the installation of the BioVizion medical
imaging system needs to meet the following requirements:
Recommended room dimensions
- 11’ total length
- 11’ total width (4’ for system installation)
- 6’ between the camera and the patient
To minimize light reflection, the lab walls must be painted a dark, flat colour.
The floor covering must also be of a non-reflective material and colour.
The doors and windows must be able to block out all exterior light and
the light switch should be located for easy access by the image technician.
Electric wall outlets (120V) must be located immediately behind or beside
both the imaging column and the patient booth.
An Internet or local network connection must be located near the
imaging column. A network connection is required in order to use
the BioVizion system as an image server. |